Meter Setup
73
Quality Control (QC) Lockout
A QC Lockout requires a quality control test to be performed. You may also set the number of
levels for this test. You may select from the following options:
No
Daily
Weekly
Monthly
All enabled QC Lockouts, if the time period has elapsed, are activated at 8:00 a.m. (presetting,
time can be changed in a PC/host system).
See page 71 on how to access the Setup-QC Settings menu.
1 Touch QC Lockout to set regularly recur-
ring quality control tests for available test
parameters.
2 Touch the button to select the time interval
of choice. Your selection is now high-
lighted.
3 Touch to save this setting, or:
Touch to exit this menu without saving
any changes.
4 Quality control tests may include one or
two levels. Select the number of levels
which must be performed (not applicable,
if you selected No).
5 Touch to save this setting, or:
Touch to exit this menu without saving
any changes.
The display automatically returns to the
previous screen.
09:15 am
QC Lockout
IQC Lockout
QC Result Format
New Lot Lockout
Reset Test Param.
Setup - QC Settings
09:15 am
No
Daily
Weekly
Monthly
Select QC Lockout