Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
203User Manual 1173.9240.02 ─ 17
Remote commands for the R&S DigIConf software
Remote commands for the R&S DigIConf software always begin with SOURce:EBOX.
Such commands are passed on from the R&S FSW to the R&S DigIConf automatically
which then configures the R&S EX-IQ-BOX via the USB connection.
All remote commands available for configuration via the R&S DigIConf software are
described in the "R&S®EX-IQ-BOX Digital Interface Module R&S®DigIConf Software
Operating Manual".
Example 1:
SOURce:EBOX:*RST
SOURce:EBOX:*IDN?
Result:
"Rohde&Schwarz,DigIConf,02.05.436 Build 47"
Example 2:
SOURce:EBOX:USER:CLOCk:REFerence:FREQuency 5MHZ
Defines the frequency value of the reference clock.
Remote commands exclusive to digital I/Q data input and output
INPut:DIQ:CDEVice....................................................................................................... 203
INPut:DIQ:RANGe[:UPPer]:AUTO................................................................................... 204
INPut:DIQ:RANGe:COUPling.......................................................................................... 205
INPut:DIQ:RANGe[:UPPer]............................................................................................. 205
INPut:DIQ:RANGe[:UPPer]:UNIT.....................................................................................205
INPut:DIQ:SRATe.......................................................................................................... 205
INPut:DIQ:SRATe:AUTO................................................................................................ 206
INPut:DIQ:CDEVice
This command queries the current configuration and the status of the digital I/Q input
from the optional Digital Baseband Interface (R&S FSW-B17).
For details see the section "Interface Status Information" for the Digital Baseband Inter-
face (R&S FSW-B17) in the R&S FSW I/Q Analyzer User Manual.
Return values:
<ConnState> Defines whether a device is connected or not.
0
No device is connected.
1
A device is connected.
<DeviceName> Device ID of the connected device
<SerialNumber> Serial number of the connected device
<PortName> Port name used by the connected device
Configuring the Measurement