R&S NRP Data acquisition and parameters (Sensor menu)
1144.1400.12 4.21 E-2
Filter dialog box:
Auto Config…
( )
ä ä
Fig. 4-18 Auto Filter dialog box
A | B | C | D ( )
ää
Tabs for selecting the sensor.
Normal –
( )
ä
Selects the autofilter Normal mode.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:TYPE[?] RESolution
Fixed Noise
( )
ä
Select the autofilter Fixed Noise mode.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:TYPE[?] RESolution
Noise Content
( )
ä ä
Selects the proportion of intrinsic noise in the measured result. Specifi-
cally,
Noise Content gives the permitted relative variation of the result
which may not be exceeded for 95% of the observation time.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:NSRatio[?] <float_value>
Max Settling Time
( )
ä ä
Specifies an upper limit for the settling time in the Fixed Noise mode.
Limit exceeded S/N
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:MTIMe[?] <float_value>
Ref Timeslot
( )
ä ä
Specifies the timeslot to which autofilter is applied when measure-
ments are being performed in the
Timeslot mode. This parameter is not
available in the other measurement modes.
Remote Control
[SENSe[1..4]]:AVERage:COUNt:AUTO:SLOT[?] <int_value>