Remote Control - Commands R&S NRP
1144.1400.12 6.80 E-3
Operation Status Register
The CONDition register contains information on the operations currently being performed by the device,
while the EVENt register contains information on the operations performed by the device since the last
readout of the register.
These two registers can be read by the queries STATus:OPERation:CONDition? or
STATus:OPERation[:EVENt]?.
Table 6-20: Meaning of bits used in the Operation Status Register
Bit
No.
Meaning
0 Operation Calibrating Status Register summary bit
This bit is set if one of the sensors is being calibrated.
1 to 3
Not used
4 Operation Measuring Status Register summary bit
This bit is set if one of the sensors is performing a measurement.
5 Operation Trigger Status Register summary bit
This bit is set if a sensor is in the WAIT_FOR_TRG state, i.e. waiting for a trigger event.
6
Not used
7 to 9
Not used
10 Operation Sense Status Register summary bit
This bit is set if a sensor is initialized.
11 Operation Lower Limit Fail Status Register
This bit is set if a displayed value has dropped below a lower limit value.
12 Operation Upper Limit Fail Status Register
This bit is set if a displayed value has exceeded an upper limit value.
13 to
14
Not used
15
Bit 15 will never be used.