R&S NRP Measuring average power (Cont Av mode)
1144.1400.12 2.7 E-2
5 Setting a fixed offset correction
– ( )
*↔ *
¾ Using the topmost rocker switch or the cursor keys
(ÕÖ), select the Sensor menu.
All sensor-related settings can be made in the
Sensor
menu. This determines the type and details of data ac-
quisition.
Offset… | Filter – ( )
*
¾ Press the Offset… softkey (left-hand side of rocker-
switch).
The Offset dialog box opens. Factors for correcting ex-
ternal signal losses or gains, due to an attenuator, for
example, can be set in this box.
Global – ( )
*
¾ Activate the editor for the global offset by pressing
the left-hand side of the rocker switch next to Global.
Positive values are for correcting losses and negative
values for gains.
1 2 3 4 5 6 7 8 9 0 . – ( )
* *
¾ If you have an attenuator at hand, enter its value and
connect the attenuator between the sensor and the
signal source, otherwise just enter 10 dB.
(¶ MENU)
*
¾ Confirm the entry.