Measuring average power (Cont Av mode) R&S NRP
1144.1400.12 2.8 E-2
Global ( )
*
¾ Activate the global offset correction by pressing the
right-hand side of the rocker switch next to Global.
(¶ MENU)
*
¾ Accept the warning concerning the triggerlevel.
Global offset correction is now on. The displayed value is
increased or decreased depending on the sign of the
offset.
Factors which are not, or only minimally, dependent on
frequency can be corrected in this way.
(¶ MENU)
*
¾ Close the dialog box.
(¶ MENU)
*
¾ Close the menu.
If you have connected an attenuator and entered its at-
tenuation, the R&S NRP will display about the same
value as before.
In the note line of the measurement window, the
symbol
indicates that global offset correction has been activated.