EasyManua.ls Logo

Rohde & Schwarz SR8000 Series - Page 135

Rohde & Schwarz SR8000 Series
199 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 5 Operation
5300.9677.71 - 5.48 - EN-4
Signal monitoring at analog inputs
Nominal input amplitude
The nominal input amplitude is the signal amplitude (500 Hz sinewave signal) that produces
100 % of the set nominal deviation at the output of the FM modulator. The analog level reg-
ulators on the input side are set to ensure 9 dB headroom on the A/D converter at this am-
plitude.
Setting item Description
Impedance Selection of the input impedance; possible settings:
600 Ohm
–>2 kOhm
Level Setting the nominal input amplitude
Value range (analog): -6 dBu through +12 dBu (default setting +6 dBu)
Value range (MPX): -1 dBu through +16 dBu (default setting +6 dBu)
Threshold Setting for the input amplitude threshold for monitoring
The setting can be entered in two ways:
as a percentage of the nominal input amplitude
in the value range: 2% through 80% (default setting 5%)
as an absolute value in dBu
in a value range dependent on the absolute level, corresponding to
input as a percentage
Display Description
On-Time If the input amplitude is continuously above the threshold during the set
time interval On-Time, the input is considered to be OK.
Value range: 1 s through 6000 s (default setting 5 s)
Off-Time If the input amplitude is continuously below the threshold during the set
time interval Off-Time, the input is considered not to be OK.
Value range: 1 s through 6000 s (default setting 5 s)

Table of Contents

Related product manuals