Chapter 4 Putting into Operation
5300.9677.72 - 4.19 - EN-4
2.3.5 Setting the Transmission Frequency
The transmission frequency is set to 97.5 MHz by default.
The "Forward power" customer test point at the transmitter output (SMA female) is set to
an output level of +7 dBm to 13 dBm at a nominal power of 97.5 MHz into 50 Ω.
The frequency response is 6 dB/octave.
1. Select Transmitter > Exciter > RF.
The RF (Radio Frequency) window opens.
2. Set the required frequency using the cursor keys and confirm with OK.
The smallest frequency step size is limited to 10 kHz by the software.
2.4 Entering Output Stage Settings
2.4.1 Setting Output Power Thresholds
1. Select Output Stage > Limits.
The Limits window opens.