Reference Manual
00809-0100-4360, Rev BA
August 2008
Rosemount 1151
A-6
Output Code M
2.0 mA dc.
Span and Zero
Output Code S
Span and zero may be accessed with local adjustments or remotely via a
HART-based communicator.
Output Codes E, G, L, and M
Span and zero are continuously adjustable.
Zero Elevation and Suppression
Output Code S, E, and G.
Zero elevation and suppression must be such that the lower range value is
greater than or equal to the (–URL) and the upper range value is less than
or equal to the (+URL). The calibrated span must be greater than or equal
to the minimum span.
Output Code L
Zero is adjustable ±10% of URL and span is adjustable from 90 to 100% of
URL.
Output Code M
Zero is adjustable ±50% of URL and span is adjustable from 50 to 100% of
URL.
Temperature Limits
Ambient
Code S: –40 to 185 °F (–40 to 85 °C).
Code E: –40 to 200 °F (–40 to 93 °C).
Code G, L, M: –20 to 200 °F (–29 to 93 °C).
Storage
Code S:–60 to 185 °F (–51 to 85 °C).
Codes E, G, L, M:–60 to 250 °F (–51 to 121 °C).