EasyManua.ls Logo

Rosemount 1151GP - Australian Certifications

Rosemount 1151GP
176 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Manual
00809-0100-4360, Rev BA
August 2008
B-3
Rosemount 1151
N1 ATEX Type n and Combustible Dust
(1151 Smart only)
Certificate No.: BAS 99ATEX3293X
ATEX marking: II 3 GD
EEx nL IIC T5 (-40°C
Ta
40°C)
EEx nL IIC T4 (-40°C
Ta
80°C)
Dust Rating: T90 °C (Ta = -20°C to 40°C)
U
i
= 45 Vdc Max
IP66
Special Conditions for Safe Use (x):
The apparatus is not capable of withstanding the 500V insulation test
required by EN 50021: 1999. This must be taken into account when
installing the apparatus.
E8 ATEX Flame-Proof
Certification Number CESI03ATEX037
ATEX Marking II 1/2 G
EEx d IIC T6 (–40 Ta 40 °C)
EEx d IIC T4 (–40 Ta 80 °C)
1180
V = 60 Vdc maximum
Australian Certifications Standards Association of Australia (SAA) Certification
E7 SAA Flame-proof
Certificate Number Ex 494X
Ex d IIB + H
2
T6
DIP T6
IP65
Special Conditions for safe use (x):
For transmitters having NPT, PG or G cable entry threads, an
appropriate flame-proof thread adaptor shall be used to facilitate
application of certified flame-proof cable glands or conduit system.
I7 SAA Intrinsically Safe
Certificate Number: Ex 122X
Ex ia IIB T5 (T
amb
= 40 °C)
Ex ia IIB T4 (T
amb
= 80 °C)
Special Conditions for Safe Use (x):
The equipment has been assessed to the entity concept and accordingly
the following electrical parameters must be taken into account during
installation.

Table of Contents