Reference Manual
00809-0100-4004, Rev BA
August 2010
Rosemount 8800D
A-18
Electrical Connections
1
/2 –14 NPT or M20 X 1.5 conduit threads; screw terminals provided for 4–20 mA. FOUNDATION
fieldbus and pulse output connections; communicator connections permanently fixed to terminal
block.
Non-Wetted Materials
Housing
Low-copper aluminum (FM Type 4X, CSA Type 4X, IP66)
Optional SST housing
Paint
Polyurethane
Cover O-rings
Buna-N
Flanges
316/316L lap joint
Temperature Sensor (MTA Option)
Type-N Thermocouple
Process-Wetted Materials
Meter Body
316L wrought stainless and CF-3M cast stainless or N06022 wrought Nickel Alloy and CW2M
cast Nickel Alloy. Other material grades available. Consult factory for other materials of
construction.
Flanges
316/316L stainless steel
Nickel Alloy N06022 Weld Neck
Collars
Nickel Alloy N06022
316/316L Stainless Steel
Process Connections
Mounts between the following flange configurations:
ASME B16.5 (ANSI): Class 150, 300, 600, 900, 1500
DIN: PN 10, 16, 25, 40, 64, 100, 160
JIS: 10K, 20K, and 40K
Weld-end: Schedule 10, Schedule 40, Schedule 80, Schedule 160
Mounting
Integral (Standard)
Electronics are mounted on meter body.
Surface Finish of Flanges and Collars
Standard: To the requirements of the applicable flange
standard.
Smooth: 63 to 125 inches
(1.6 to 3.1 meters) Ra roughness