5 - 16 HS70A Service Manual
Electric Shocks
The probe uses electrical energy. If it touches conductive materials, there are risks of electric
shocks to the patient or the user.
WARNING
Have the system regularly inspected by Samsung Medison technicians for
electric leaks.
Do not immerse the probe into liquid.
Do not drop the probe or apply mechanical shocks.
Inspect the housing, strain relief, lens and seal for damage, and check for
any functional problem before and after each use.
Do not apply excessive force to twist, pull or bend the probe cable.
The power protection fuse protects the probe and the product from excess
current. If the power monitoring protection circuit detects excess current, it
immediately shuts off the current to the probe in order to prevent the probe
surface from overheating and to restrict the ultrasound power output.
The temperature of the product for making contact with patients is limited
under 43°C. The ultrasound power output (AP&I) is in compliance with US
FDA standards.