Chapter 3. Installing the Product 3 - 5
Installing the Product
Installation Safety
In a hospital environment, dangerous electric current may occur as a result of the potential
difference between a contactable conductive part and connected equipment in treatment rooms.
The solution to the problem is consistent equipotential bonding. Medical equipment is connected
with connecting leads made up of sockets which are angled to the equipotential bonding network
in medical rooms.
[Figure 3. 2 Equipotential bonding]
CAUTION
If the product needs to be transported or stored for an
extended duration, the temperature and humidity of the
environment must be checked.
A sudden change in temperature may cause condensation and
lead to product failure.
Refer to "[Table 3-2. Operational Temperature of Product]" before turning the product on.
Temperature
-20 -15 -10 -5 0 5 10 ~ 35 45 50 55 60
Waiting time
Hour
16 10 8 6 4 2
Use
immediately
2 4 6 10
[Table 3-2. Operational Temperature of Product]