5 - 42 HS70A Service Manual
Assembling the Biopsy Kit
Disposable Biopsy KIT
1. Place a Sheath up to the top of the probe’s handle.
2. Mount the biopsy adapter onto the probe. If the surface of probe is fluted, mount the
adapter in accordance with it.
3. Install the needle into the needle guide and start the exam.
Reusable Biopsy KIT
1. Mount the biopsy adapter onto the probe.
2. Insert gel into the sheath. And cover the probe completely.
3. Install the needle guide onto the adapter.