5 - 46 HS70A Service Manual
Cleaning and Disinfecting the Biopsy Kit
WARNING
Always use protective eyewear and gloves when cleaning and disinfecting the
biopsy kit.
Wash and disinfect the biopsy kit to reduce pathogens to the level of 10
-6
. Some components of
the biopsy kit may be disposable. Please read the biopsy kit user manual carefully before use.
Cleaning and Disinfecting a Stainless Steel Biopsy Kit
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its component parts, if applicable.
3. Using a small brush and water, scrub each part to remove trapped material from the
biopsy kit.
4. Rinse with water to remove remaining particulates.
Disinfection
1. Disinfect the adapter by autoclaving (Steam) or using gas (Ethylene Oxide).
2. After disinfection, follow the proper post-disinfection procedure for the disinfection
method used. (Please refer to the disinfection user manual, etc.)
3. Inspect the components for damage such as cracks, rust or breakage. If damage is
evident, discontinue use of the biopsy kit and contact Samsung Medison’s servicing
department.
Cleaning and Disinfecting a Plastic Biopsy Kit
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its component parts, if applicable. Discard the single-
use parts. These parts cannot be disinfected.
3. Using a small brush and water, scrub each part to remove trapped material from the
reusable components.
4. Rinse with water to remove remaining particulates.