Initial Startup
Operating Instructions Arium
®
Comfort I, Comfort II 29
Ultrapure Water Cartridge Installation
t Remove the new cartridge from the packaging and write the date of
installation under “Date of installation” on the label.
t To facilitate the connection of the adapter to the cartridge, don gloves to
moisten the O-rings (1) on the connection adapter with distilled water.
Avoid direct contact with skin at the connections in order to prevent any
microbial contamination.
t Open the safety bars (2) of the connection adapter. Next, press the
connection adapter (3) firmly into place on the cover of the cartridge.
Push the connection adapter so far down that the locking tabs are
located at the level of the spacers.
t Push both locking tabs all the way under the upper ring of the spacers.
t Insert the cartridge into the housing. The label must be facing you.
t Close the door on the unit.