Operation
Operating Instructions Arium
®
Comfort I, Comfort II 37
7.2.3 Time-Controlled Dispensing
t To perform time-controlled dispensing, place an appropriately large
vessel under the sampling outlet and in the operating mode press the
(R) “ ” key.
t The currently set dispensing time is shown in the display. To change the
dispensing time, select Set dispensing time.
t Set the desired dispensing time.
The following dispensing times can be adjusted according to the given scale
intervals. The changed dispensing time (2.5 min in the example) is indicated
in the display.
Dispensing time Scale interval
0.5 min – 10 min 0.5 min
10 min – 30 min 1 min
30 min – 60 min 5 min
The adjustable dispensing time depends on the connected Arium
®
Bagtank.
t Switch to “Time-controlled” and start volume-controlled dispensing.
y The remaining dispensing time is indicated in the display.
t Time-controlled dispensing can be canceled by pressing the function
key (R) “
Stop
”. The system then switches back to the operating mode.
The termination of time-controlled dispensing is signaled acoustically via
the confirmation beep (see Chapter "8.3.13 Acoustic Signals", page 50).
Dispense
Set favorites
Volume-controlled
0.2 l
Time-controlled
Dispense
Set dispensing time
2.5 min
Dispense
Volume-controlled
0.1 l
Time-controlled
2.5 min
Start
11:43 10.04.15
0.055 µS/cm c
1.0 ppb TOC
2.0 min
Stop