System Menu
Operating Instructions Arium
®
Comfort I, Comfort II 41
8.2 Save Data
The save data functions are used for output of current measured values and
system information to a printer or to save data on an SD memory card.
Access the menu item Menu Save data. Three different options are
available for printing and saving: Single printout, Print interval and Print
when dispensing.
Printing and saving can be used at the same time as when Print interval and
Print when dispensing are activated. The checkboxes all show active Save
data functions.
Printer
For data printing, a data printer of the type Sartorius YDP30 should be used.
SD Card
For digital data storage, an SD memory card can be inserted in the Arium
®
Comfort system (right side, top). SD cards are supported with a max.
capacity of 4 GB. The data are saved in the Excel-compatible csv format.
With the Arium
®
Comfort system, it is not possible to delete saved data and
| or format the SD card.
The following symbols appear in the first line of the display whenever
printing and | or saving are taking place:
Print symbol:
Disk symbol:
8.2.1 Parameter Selection
Under Menu Save data Settings select the parameters to be printed
out and | or to be saved.
Using the checkbox, the following parameters can be selected and
deselected:
System Information
LFP (conductivity of the ultrapure water)
TOC (only for system with UV lamp & TOC)
Temperature (ultrapure water)
LFA (conductivity according to pre-stage, on Comfort II version only)
LFR (conductivity according to pre-stage, on Comfort I version only)
LFF (conductivity of the feed water) (compensated)
Menu r Save data
Single printout
Print interval
Print when dispensing
Save once
Start