Operation
Operating Instructions Arium
®
Comfort I, Comfort II 39
7.2.5 Ultrapure Water Dispensing through the Sampling
Tubing
Water can also be dispensed through the sampling tubing included with the
equipment supplied. It may be advisable to use the outlet tubing when, for
example, you need to fill a larger vessel.
To connect the tubing, proceed as follows:
t Release the final filter from the quick connector on the display |
dispenser unit by simultaneously pulling out the filter and pressing the
retaining ring on the component.
t Press the sampling tubing onto the quick connector of the display |
dispenser unit.
t Attach the straight connector to the free end of the sampling tubing.
t Press the final filter into the straight connector.
t Allow 6 liters of water to run through the final filter (e.g. using manual
and | or volume-controlled water dispensing); this is done to rinse the
final filter.
t Vent the final filter using the attached vent valve.
t After dispensing ultrapure water, attach the protective cap to the bell
assembly.
The rinsing of the final filter can be omitted if no new final filter is used.