Operation
34 Operating Instructions Arium
®
Comfort I, Comfort II
7 Operation
Once initial startup is complete, your Arium
®
Comfort system will be in the
operating mode. The pre-stage continuously fills the Arium
®
Bagtank until it
is completely filled. At the same time, you can dispense ultrapure water.
7.1 Operating Mode
If the connected Arium
®
Bagtank is completely full, backflushing takes place
for 4 minutes to rinse the RO modules with ultrapure water directly from the
Bagtank. During backflushing, you can continue to dispense ultrapure
water.
Once the backflushing process is complete, the device will revert to
operating mode. The filling level of the tank appears on the display again.
Once the tank level drops to 75%, pure water production begins until the
Arium
®
Bagtank has filled up again completely.
It is recommended that you do not turn your Arium
®
Comfort system off
during normal operation (e.g., evenings and weekends) to ensure uniformly
high ultrapure product water quality. The Eco mode is available for this.
If the tank volume falls below 15%, the Arium
®
Comfort system issues an
alert on the display. An empty tank can damage the system. Therefore, when
dispensing directly from the tank (pure water), always make sure that the
tank is sufficiently full (tank fill level in the display).