Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 79
9.13 Depressurization
It is necessary to depressurize the system when replacing the UV lamp, for
example. Additionally, Sartorius recommends you depressurize your Arium
®
Comfort system prior to transport. In the latter case, the step “Restart after
depressurization” should be carried out at the new place of operation.
The depressurization routine is activated in the menu item Care
Depressurization.
t Disconnect the “Tank Inlet” tubing from your Arium
®
Comfort system.
t Place a vessel (1l) under the water outlet and start depressurization.
y The water pressure in the system is reduced. Water flows from the
sampling outlet.
t Switch off your Arium
®
Comfort system at the main switch.
This message can only be cancelled by switching off the device!
Carry out the pending work (e.g. replace the UV lamp).
Care r Depressurization
Disconnect “Tank Inlet” tubing
from device
Cancel OK
Care r Depressurization
Collect water leaking from
dispenser unit in a vessel (1l)
and start
Start
Care r Depressurization
Depressurization
0.5 min
Cancel
Care r Depressurization
Switch off unit at main switch