Care and Maintenance
64 Operating Instructions Arium
®
Comfort I, Comfort II
t Remove the old prefilter by unscrewing the upper gray (3) and lower red
(4) tubing connections.
t Remove the new prefilter from the packaging and write the date of
installation under “Date of installation” on the label.
t Connect the new prefilter. Connect the upper gray (3) and the lower red
(4) tubing connectors.
t Press the new prefilter into the holder (2).
t Insert the ultrapure water cartridge (1) back in the device and close the
device door.
t Start the rinsing process.
y The new prefilter and | or the softener are rinsed for 30 minutes.
t Connect the “Tank Outlet” tubing to the Arium
®
Bagtank.
Replacing the prefilter and | or the softener is complete. The system
switches back to operating mode.
The prefilter timer is reset automatically.
3
4
Care r Prefilter
Start rinsing process
Start
Care r Prefilter
Rinsing
30 min
Cancel
Care r Prefilter
Connect “Tank outlet” tubing
to tank
OK