Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 67
t Start the rinsing process.
y The new RO modules and | or the prefilter and | or softener are rinsed for
60 minutes.
t Connect the “Tank Outlet” tubing to the Arium
®
Bagtank (“IN” tank
connection).
Replacing the RO modules is complete. The system switches to operating
mode.
The timer for the RO module and | or RO modules is reset automatically.
Care r RO Modules
Start rinsing process
Start
Care r RO Modules
Rinsing
60 min
Cancel
Care r RO Modules
Connect “Tank outlet” tubing
to tank
OK