Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 69
t Open the door on your Arium
®
Comfort systems and push the ultrapure
cartridge (1) to the side.
t Remove the softener (2) from the holder.
t Remove the old softener by unscrewing the upper blue (3) and lower
gray (4) tubing connections.
t Unpack the new softener and write the date of installation of the
softener under “Date of Installation” on the label.
t Connect the new softener. Connect the upper blue (3) and the lower
gray (4) tubing connectors.
t Press the new softener into the holder (2).
t Insert the ultrapure water cartridge (1) back in the device and close the
device door.
1
2
3
4