Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 73
t Unscrew the cover cap by turning counterclockwise (3) and push it back
along the cable (4).
t Carefully unplug the adapter from the UV lamp (5).
t Carefully remove the old UV lamp unit from the stainless steel
t Carefully insert the new UV lamp into the stainless steel chamber.
NOTICE
Never touch the glass body with your fingers. Fingerprints will reduce the
service life of the lamp. If you touch the lamp, clean it with a damp, lint-free
cloth. If required, use isopropanol for cleaning.
t Carefully insert the adapter onto the new UV lamp.
t Make sure that the adapter is inserted at the proper orientation.
t Guide the plastic cover over the lamp socket and carefully screw it into
the stainless steel chamber (clockwise), hand-tighten.
t Remove the collection vessel.
t Push the excess cord of the connection cable back inside the device.
t Insert the cartridge back in the device and close the device door.
t Turn on the Arium
®
Comfort system at the power switch.
t Reset the UV lamp timer under Care UV lamp timer.
3 4
6 5
Care Timer UV lamp
Reset UV lamp timer
Cancel