43
Configuring Weighing Platforms
t Press J to apply linearization weight 1 or press ( to cancel the linearization function.
y After the linearization weight 1 has been saved you will be prompted to place the second
linearization weight on the weighing pan.
t Repeat the procedure for all required linearization weights.
y After the last linearization weight has been saved you will be prompted to remove any load
from the weighing pan.
t Remove all linearization weights from the weighing platform.
y The zero point is applied automatically after a brief time. The indicator automatically
switches to weighing operation.
When linearization is complete, the GMP-compliant printout shown on the left is generated.
t Slide the menu access switch to the right (= “closed" position).