External Calibration and Manual Adjustment with Default
Weights (weighing parameters: factory settings)
Preparation
t Select Setup: Press the M key.
t Select device parameters: Press the “O" soft key.
t Select weighing platform “WP 1," WP 2:" Press the “O" soft key or
t Select interface “COM 1," COM 2"orUNICOM" (depending on the interface):
Press the “
O" soft key.
t Select weighing platform WP 3: Press the “O" soft key.
Calibration/Adjustment
CAL key function
o Ext. cal./adj.; factory-def. wt.
Ext. cal./adj.; user-defined wt.
Key blocked
Cal./adj. sequence
Cal. then auto adj.
o Cal. then manual adj.
isoCAL function (for Sartorius IS weighing platforms only)
oOff
Adjustment prompt
Activate ext. adj.
o Activated
Deactivated
Parameter for external weight
o = Factory setting
t To save settings and exit the Setup menu: Press the M key or the “oo" soft key.
Procedure
t Zeroing the scale by pressing the ( key.
t Start calibration/adjustment using the J key.
y The display “C.EXT.D" appears for two seconds.
y You are prompted to place the required weight on the platform (e.g., 10,000 g).
Operation
C.EXT.D
C.EXT.DEF.
12