Operation
Features
Which of the following features are available depends on the connected weighing
platform. These features are configured in the Setup menu:
External calibration/adjustment blocked in verified weighing instruments
- External calibration/adjustment with the default weight value or user-defined weight
(not available on verified instruments) under
“…
Calibration/adjustment:CAL key function"
- Specify the weight for external calibration/adjustment under
“…
Calibration/adjustment:Parameter for external
weight"
- Internal calibration/adjustment of IS weighing platforms
(configuration in “
COM 1:, COM 2: or UNICOM:WP 3 )
- Block the J key to prevent the use of the two functions described above under
“…
Calibration/adjustment:CAL key function"
- Calibrate first; then adjust automatically or manually (not for verified weighing
instruments) under “…
Calibration/adjustment:Cal/adj.
sequence"
- For Sartorius IS weighing platforms only: Flashing WP symbol as adjustment prompt
(If more than one weighing platform is connected, the platform number is also displayed)
under
“…
Calibration/adjustment:CAL key function"
- Block/Unblock external calibration/adjustment under
“…
Calibration/adjustment:Activate ext. adj."
- Altitude and latitude or gravitational acceleration displayed after “
Cal" is displayed at
the start of the calibration/adjustment process if these values are supported by the
weighing platform under “
Device parameters:Operating
parameters:Display geogr. data"
For each of these parameters, the term is displayed first (Altitude, Latitude or Gravity)
for 1 second, and then the corresponding value is displayed continuously until you press
J.