69
Weighing
This application is always available during operation.
Features
– Zeroing by pressing (
– Storing the weight on the platform as a tare by pressing )
– Tare container weight automatically
– Use a barcode scanner to enter tare weight
– Use the numeric keys 0 to enter a tare weight and press ) to save
– Delete tare values using the c key and save using the ) key
– Toggle the display using the k key between:
– 1st and 2nd weight unit
– SQmin
– You can configure the k key function in the “
Fn key" Setup menu
– 10-fold increased resolution using the K key
– Weighing using up to three weighing platforms
– Individual numeric ID codes for weight values
– Print weight values:
– Manually, by pressing p
– GMP printout (see “Data Interfaces")
– Automatic printout
– Automatic data output (see “Data Interfaces")
– Restore factory settings in the Setup menu:
"
Application parameters:Factory settings: only
application:Yes"
Soft Key Functions
ID Enter up to six ID codes for identifying results on the printout
1st ID Save the value entered as the first ID code.
Preparation
t Select Setup: Press M.
t Select Application parameters: Press the “O" soft key.
t Select “Application 1 (basic settings)":
Press the “
O" soft key.
t Confirm the “Weighing" application: Press the “l" and “o" soft keys.
t Select the “Min. load f. auto. taring/printout" function:
Press the “
O" soft key.
t Confirm the “o 10 digits" setting:
Press the “
l" and “o" soft keys.
t Select the “Factory setting" menu item: Press the “O" soft key.
t Confirm the “o No" setting: Press the “l" and “o" soft keys.
t To save settings and exit the Setup menu: Press the M key or the “oo" soft key.
Automatic Taring
The first weight on the scale that exceeds the preset minimum capacity is stored in the
tare memory at stability. The values for subsequent loads are stored as weight values.
The scale returns to the initial state when the load on the scale is less than 50% of the
minimum load. Operating menu setting: “
Application parameters:
Autotare 1st weight:Yes"
Operation