EasyManua.ls Logo

Sartorius Combics 3 Series - Sqmin Operation

Sartorius Combics 3 Series
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
SQmin Operation
Configuration in Service Mode
The SQmin value can only be entered in Service mode.
t Remove the cap.
t Slide the menu access switch to the left (= “open" position).
If the device is part of a verified weighing facility, this will only be possible if the
verification seal is broken. The weighing equipment must then be verified again.
t Activate the Service mode, see “Service Mode."
t In the “Device parameters" menu, select “SQmin" and open using the
O" soft key.
t Select “SQmin input" using the O" soft key.
t Select SQmin WP1" using the O" soft key, enter the value via the keypad (in this
example 0.100 kg) and save using the
l" soft key or cancel using “ESC."
t Then enter the values for WP 2 and WP 3.
t Press M or “oo" to exit the Setup menu.
t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The device is now in normal weighing mode.
Configuration in Weighing Mode
t In the “Device parameters" menu, select “SQmin" and open using the
O" soft key.
t Select “Display.".
t Open using the “O" soft key.
t Confirm the factory setting “oYes" using the “l" soft key.
t Press the “o" soft key several times to switch to the next highest menu.
The SQmin display must be turned on to use the SQmin function.
t In the “Fn key:SQmin display:" menu, select using the “O" soft key and
confirm using the “
l" soft key.
t Press M or “oo" to exit the Setup menu.
84