PR 5211 Instrument Manual PR 5211 Transmitter Series
Sartorius EN-9
2.2 Transmitter Survey
- Accuracy 6000 d @ 6 samples/sec
- Internal resolution 4.8 Mio counts
- Linearity <0.002% (typ.)
- Sample rate selectable: 6 ... 100/sec
- Digital filter with selectable characteristic
- Interfaces are galvanically isolated
- 3 pairs of programmable limits
- 24 V DC Supply voltage
- Connection using 7 plug-in terminal blocks
- Connector (female) for ProfiBus
- Connector USB-B for PC
- Connection cable (1.8 m) for PC (USB 2.0; A>B)
- The unit is suitable for snap-in mounting (rail).
- 5 status LEDs for power, communication, error detection
- Calibration and configuration of the instrument are menu-
guided via PC or via ProfiBus commands.
- Calibration by means of weights or by entry of the mV/V
values without additional calculations
- Analog Output 0/4…20 mA, configurable for gross/net (PR
5211/00 and PR 5211/10 only)
- Analog Output value defined by ProfiBus (PR 5211/00 only)
- Digital Input 3 channels, galvanically isolated
- Digital Output 3 channels, galvanically isolated
Communication protocols
For the internal RS-485:
- Remote display protocol
- SMA protocol
Slave:
- PR 5211/00 ProfiBus-DP
- PR 5211/11 ProfiBus-DP

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5211 Series and is the answer not in the manual?

Sartorius PR 5211 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5211 Series
CategoryTransmitter
LanguageEnglish

Related product manuals