PR 5211 Instrument Manual ProfiBus Interface
Sartorius EN-69
6.2.15 Register 112125: Action bits transition controlled (write)
Set bits, see chapter 6.1.5.3
As soon as the bit is set, it is reset internally and the action is executed, it is transition triggered (for writing
once).
The bit is set with the shown number as Write_Value_Select (see chapter 6.1.5.3).
Register 112 SetZero
Register 113 SetTare
Register 114 ResetTare
Register 115 SetTest
Register 116 ResetTest
Register 117 ResetPwf
Register 118 SetFixTare
Register 119 GetFixTare
Register 121 ResetError
Register 122 SaveProcess
Register 123 SetParam
Register 124 GetParam
Register 125 SaveConfig
To prevent the EAROM from being written too often, the writing rate for taring and zero setting should not be
shorter than 15 seconds, for configuration data not shorter than 5 minutes.