PR 5211 Instrument Manual Commissioning
Sartorius EN-25
4.1.2 Factory Settings
Calibration data <default> Calibration parameters <default>
Full scale (FSD) <3000> <Kg> Measuring time (M) <320>ms
Step width <1> Measuring rate <160>ms
Deadload <0.000000>mV/V Standstill time <1>M
Span <1.000000>mV/V Standstill range <1.00>d
Standstill timeout <8>M
Calibration parameters <default>
Test mode <absolute>
Overload (range above FSD) <9>d Zeroset range <50.00>d
* W
&
M mode <off> Zerotrack range <0.25>d
Filter <off> Zerotrack step <0.25>d
Frequency <1.56 Hz> Zerotrack repeat <0>M
* Default: ‘off’
At delivery the calibration data and parameters are set to default (factory settings).
If a new calibration is started, the calibration data are set to default (The parameters remain unchanged).
4.1.3 Power Failure
Calibration data and parameters are stored in EAROM and kept in case of power failure or if the instrument
is disconnected from power supply.