EasyManuals Logo

Sartorius PR 5211 Series User Manual

Sartorius PR 5211 Series
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #48 background imageLoading...
Page #48 background image
SMA Protocol PR 5211 Instrument Manual
EN-46 Sartorius
4.13.1 Analog Part/Weight Status
Display Error 3 Error 7 negative
0
positive
FSD
(SKE)
Error 2 Error 3
±0,25d
overload
Bit
E3
E7
Below0
CZERO
aboveFSD
E2
E3
InZSR
DeadLd±InsideZSR
ADUERR
ADUERR
ADUERR
W&M off
-
-
-
DIMM
5 SMA Protocol
5.1 General
The Scale Manufacturers Association (SMA) protocol provides a simple access to the scale. Data could be
read and functions could be executed.
The RS-485 interface is used. The setting of the interface is fixed to 8 bit, no parity and 1 stop bit.
The scale commands are printable ASCII characters and start with a <LF> (0A hex) and end with a
<CR> (0D hex).
On each received command the transmitter sends after about 100ms an answer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5211 Series and is the answer not in the manual?

Sartorius PR 5211 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5211 Series
CategoryTransmitter
LanguageEnglish

Related product manuals