Technical Data  PR 5211 Instrument Manual 
EN-86    Sartorius 
10.4.2  Electromagnetic Compatibility (EMC) 
All data comply to NAMUR NE 21 and EN 61326 
Housing  Radio frequency electromagnetic 
field (80 – 3000 MHz) 
EN 61000-4-3  10 V/m 
Electrostatic discharge (ESD)  EN 61000-4-2  6 kV/8 kV 
       
Signal and control lines  Electrical fast transients (Burst)  EN 61000-4-4  2 kV 
Peak voltage (surge) 1.2/50 µs  EN 61000-4-5  2 kV 
Conducted disturbances by radio-
frequency (0.15 – 80 MHz) 
EN 61000-4-6  10 V 
       
DC input power port  Electrical fast transients (Burst)  EN 61000-4-4  2 kV 
Peak voltage (surge) 1.2/50 µs  EN 61000-4-5  1 kV/2 kV 
Conducted disturbances by radio-
frequency (0.15 – 80 MHz) 
EN 61000-4-6  10 V 
Voltage variations  NAMUR NE21  100%-0%-100%,  
2 s-1 s-2 s 
100%-40%-100%,  
2 s-1 s-2 s 
Voltage interrupts  NAMUR NE21  0%, 20 ms 
10.4.3  RF interference suppression 
acc. To EN 55011 group 1, limit value class A