Safety Hints PR 5211 Instrument Manual
EN-6 Sartorius
1.4.1 Installation
The instrument is designed for mounting on standard rails (35 mm, acc. to DIN 46277).
Excessive heat may reduce the instrument lifetime!
When mounting on the rail, make sure that the distance from other instruments left and
right of the module is at least 20 mm.
1.4.2 Electrostatically Sensitive Components
This instrument contains electrostatically sensitive components. Therefore, potential equalization must be
provided when working at the instrument (antistatic protection).
1.4.3 Protective Earth
The connection to protective earth is done via the mounting rail.
1.4.4 Supply Voltage Connection
The supply voltage is 24 V DC +10 %/-15 %.
The power consumption is 8.2 W maximum.
For connection to 230/115 V AC an external power supply (e.g. Sartorius PR 1624/00
or Phoenix Mini Power) is required.