EasyManuals Logo

Sartorius PR 5211 Series User Manual

Sartorius PR 5211 Series
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
SMA Protocol PR 5211 Instrument Manual
EN-50 Sartorius
5.3.11 About Scale Scroll
Command: <LF>B<CR>
Response: The scale will send the rest of the ABout the scale data.
<LF><MFG>:<yyyyyy><CR>
For detail see chapter 5.4.5.
5.3.12 Scale Information
Command: <LF>I<CR>
Response: The scale will send the first line of the scale Information data.
<LF><SMA>:<yyyyyy><CR>
For detail see chapter 5.4.6.
5.3.13 Scale Information Scroll
Command: <LF>N<CR>
Response: The scale will send the rest of the scale INformation data.
<LF><TYP>:<yyyyyy><CR>
For detail see chapter 5.4.6.
5.3.14 Abort Command
Command: <ESC>
Response: This is the only command that the scale receives which does not follow the <LF>c<CR>
protocol and does not have a response. The <ESC> character can be detected at any time
and any command is aborted.
5.3.15 Repeat Displayed Weight Continuously
This is a command with Unsolicited Response. It defines a host command that is not considered strictly
command/ response. This is because the scale will respond continuously as commanded to do so by the host.
Command: <LF>R<CR>
Response: Scale Repeats weight and status information continuously until another command is
received.
<LF><s><r><n><m><f><xxxxxx.xxx><uuu><CR>
For detail see chapter 5.4.1.
Depending on the used baud rate, the following approximately repetition times of response messages are
possible:
19200 Bd
100 ms
9600 Bd
110 ms
4800 Bd
170 ms

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5211 Series and is the answer not in the manual?

Sartorius PR 5211 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5211 Series
CategoryTransmitter
LanguageEnglish

Related product manuals