PR 5211 Instrument Manual  SMA Protocol 
Sartorius    EN-49 
5.3.5  Request Scale to Tare 
Command:  <LF>T<CR> 
Response:  Scale attempts to tare itself using the weight on the scale display and reports the tare 
status in the <s> and <n> status indicators. 
  For this function the standstill condition of the scale must be fullfilled. The maximum wait 
time for standstill could set with the 'Standstill time-out', see chapter 4.9.2.8. 
  <LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR> 
  For detail see chapter 5.4.1. 
5.3.6  Set Scale Tare Weight 
Command:  <LF>T<xxxxxx.xxx><CR> 
Response:  Scale attempts to take the <xxxxxx.xxx> data as the tare weight and reports the tare 
status in the <s> and <n> status indicators. 
  <LF><s><r><N><m><f><xxxxxx.xxx><uuu><CR> 
  For detail see chapter 5.4.1. 
5.3.7  Return Tare Weight 
Command:  <LF>M<CR> 
Response:  Scale returns tare weight stored in scale Memory and identifies the weight as tare in the 
<n> status indicator. 
  <LF><s><r><T><m><f><xxxxxx.xxx><uuu><CR> 
  For detail see chapter 5.4.1. 
5.3.8  Clear Scale Tare Weight 
Command:  <LF>C<CR> 
Response:  Scale Clears tare weight and reports the tare status in the <n> status indicator. The tare of 
scale is reset. 
  <LF><s><r><G><m><f><xxxxxx.xxx><uuu><CR> 
  For detail see chapter 5.4.1. 
5.3.9  Invoke Scale Diagnostics 
Command:  <LF>D<CR> 
Response:  The scale runs scale diagnostics and sends a diagnostic response message with the results 
of the tests. 
  <LF><r><e><c><m><CR> 
  For detail see chapter 5.4.4. 
5.3.10  About Scale First Line 
Command:  <LF>A<CR> 
Response:  The scale will send the first line of the About scale data. 
  <LF><SMA>:<yyyyyy><CR> 
  For detail see chapter 5.4.5.