Commissioning PR 5211 Instrument Manual
EN-34 Sartorius
4.9.2.6 Standstill Range
Permissible range: 0.0050.00 d, default: 1.00 d
4.9.2.7 Test Mode
Determination, whether the test measurement displays FullScale (absolute) or the deviation related to
FullScale (relative) is made. Example: FSD = 3000, result: Should be 3000 for absolute, should be 0 for
relative.
Calibration (with/ without weights) is completed with a test measurement and the result is scaled so that
FullScale is displayed.
Default: ‘absolute
4.9.2.8 Standstill Time-out
Unless a tare- or Zero-set-command can be executed within time n (n = multiples of measuring times), e.g.
because the scale does not fulfill the standstill condition, the transmitter generates a message (e.g. no
standstill) and the command is aborted.
Enter the time in multiples of measuring time, 1…100, default: 8.
4.9.2.9 Zero Set Range
Definition of a +/- range around the calibration zero, within which
the displayed gross weight can be set to zero (by a corresponding external command), or
automatic zero tracking is active, see chapter 4.9.2.12.
Permissible range: 0.00500.00 d, default: 50.00 d
4.9.2.10 Zero Track Range
This function is only valid, if Zerotrack repeat is not set to 0 (see chapter 4.9.2.12)!
The zero tracking does only work as long as the weight signal is still in the zero set range.
Permissible range: 0.00…500.00 d, default: 0.25 d
4.9.2.11 Zero Track Step
The automatic zero tracking stepwidth must be smaller than the standstill range.
This function is only valid, if Zerotrack repeat is not set to 0 (see chapter 4.9.2.12)!
Permissible range 0.00…10.00d, default: 0.25 d
4.9.2.12 Zero Track Repeat
With the scale in standstill condition and the gross weight within the zero set range (Zeroset range),
automatic zero tracking is stepwisely done (Zerotrack step) at defined intervals (Zerotrack repeat).
Entry is in multiples of measuring time, permissible range 0 to 100, default = 0
(Automatic zero tracking = off)
Switching off the automatic zero tracking is by setting Zerotrack repeat to = 0.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5211 Series and is the answer not in the manual?

Sartorius PR 5211 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5211 Series
CategoryTransmitter
LanguageEnglish

Related product manuals