ProfiBus Interface PR 5211 Instrument Manual
EN-66 Sartorius
6.2.6 Register 5: Transmitter Type and Version
Only reading is allowed.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
TYPE MSB
Byte 1
TYPE LSB
Byte 2
MAINVERSION
Byte 3 SUBVERSION
e.g. 5210 rel. 1.23 = 52100123
hex
6.2.7 Register 6: Board Number
Only reading is allowed.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Board number MSB
Byte 1
""
Byte 2
""
Byte 3
Board number LSB
e.g. 148388723 = 08D83B73
hex
6.2.8 Register 7: (Reserved)
6.2.9 Register 8 ...14: Weight data
Only reading is allowed.
Gross, net, tare
Are stored as DINT-fixed point. The real data value is derived from DINT and EXPO as follows:
Value
Real
=Readout
DINT
*10
(-EXPO
Register 8 Actual gross value
Register 9 Actual net value, if tared, else gross
Register10 Actual tare value, if tared, else 0
Register11 Reserved
Register12 Reserved
Register13 Reserved
Register14 Full scale deflection FSD
Register15 Reserved (free)