ProfiBus Interface PR 5211 Instrument Manual
EN-66 Sartorius
6.2.6 Register 5: Transmitter Type and Version
Only reading is allowed.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
TYPE MSB
Byte 1
TYPE LSB
Byte 2
MAINVERSION
Byte 3 SUBVERSION
e.g. 5210 rel. 1.23 = 52100123
hex
6.2.7 Register 6: Board Number
Only reading is allowed.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
Byte 0
Board number MSB
Byte 1
""
Byte 2
""
Byte 3
Board number LSB
e.g. 148388723 = 08D83B73
hex
6.2.8 Register 7: (Reserved)
6.2.9 Register 8 ...14: Weight data
Only reading is allowed.
Gross, net, tare
Are stored as DINT-fixed point. The real data value is derived from DINT and EXPO as follows:
Value
Real
=Readout
DINT
*10
(-EXPO
Register 8 Actual gross value
Register 9 Actual net value, if tared, else gross
Register10 Actual tare value, if tared, else 0
Register11 Reserved
Register12 Reserved
Register13 Reserved
Register14 Full scale deflection FSD
Register15 Reserved (free)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5211 Series and is the answer not in the manual?

Sartorius PR 5211 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5211 Series
CategoryTransmitter
LanguageEnglish

Related product manuals