ProfiBus Interface PR 5211 Instrument Manual
EN-80 Sartorius
6.3.3.20 Parameter P53: A/D Converter Sample Time (Measuring Rate)
Read only
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 SAMPLETIME
SAMPLETIME 0: 6.25 Hz == 160 ms = default
1: 12.5 Hz == 80 ms
2: 25 Hz == 40 ms
3: 50 Hz == 20 ms
4: 100 Hz == 10 ms
6.3.4 Parameter P99: Access Code
Write only
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 CODE MSB
Byte 1 ""
Byte 2 ""
Byte 3 CODE LSB
CODE If an access code was entered in the configuration tool, parameters and the registers for limits can
only be written if the access code is contained in register P99. Parameter P99 can always be written.
To activate the protection P99 has to be set to –1 again. If CODE is set to 0, no access check is done.