PR 5211 Instrument Manual ProfiBus Interface
Sartorius EN-57
6.1.4 Description of I/O Area (Read/Write Window)
6.1.4.1 Output Area
Data are transferred from PR 5211 to the master via the output area. The PR 5211 has writing access, the
master has reading access.
The weight value is transferred as 32bit binary number with sign.
Datatype: DINT
Acknowledgement of transferred value.
The function selected with Write_Value_Select is executed once.
This bit is erased if Write_Value_Select is set to 0.
Will be set at power on of the transmitter.
Will be reset at 0
1 transition of ResPower.
The transmitter is busy with executing a command. (e.g. the transmitter got
a taring command and is waiting for StandStill)
The transmitter has interrupted the execution of a command (e.g. within the
defined StandstillTimeout duration StandStill could not be reached) The
error number can be read at Lasterror. It is only set if an action is executed!
The transmitter has been tared.
The transmitter is/has been configurated. If this bit is 1, the scale
parameters (Expo/Unit/Step) have to be read again. Will be set after Power
on and reset after reading of FSD .
The transmitter is executing the ADU-Test. The weight value read is not the
gross value but the testvalue.
The transmitter is in stand still
The weight value is within zero set range
The weight value is within center zero ( ABS(Gross)<=0,25d )
The weight value is negative (Gross< -0,25d)
The weight value has exceeded the measuring range, no valid weight data
are given (Gross>FSD+Overload)
The weight value has exceeded FSD, but is still within FSD+Overload.
Error in AD conversion.(Details are available at register 1,