EasyManua.ls Logo

Sartorius PR 5211 Series - Page 81

Sartorius PR 5211 Series
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PR 5211 Instrument Manual ProfiBus Interface
Sartorius EN-79
6.3.3.16 Parameter P48: Zerotrack Range
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 RANGE MSB
Byte 3 RANGE LSB
RANGE 0 ... 500.00 d default = 0.25 d
Value = d*100
6.3.3.17 Parameter P49: Zerotrack Step
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 STEP MSB
Byte 3 STEP LSB
STEP 0 ... 10.00 d default = 0.25 d
Value = d*100
6.3.3.18 Parameter P50: Zerotrack Repeat
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 REPEAT
REPEAT 0 ... 100 measuring times default = 0 (Zero tracking is off)
6.3.3.19 Parameter P51: Overload
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0 OVL MSB
Byte 1 ""
Byte 2 ""
Byte 3 OVL LSB
OVL Overload in d 0…9999999 default = 9 d

Table of Contents

Related product manuals