EasyManuals Logo

Sartorius PR 5211 Series User Manual

Sartorius PR 5211 Series
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
ProfiBus Interface PR 5211 Instrument Manual
EN-78 Sartorius
6.3.3.12 Parameter P44: Standstill Time
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 STIME
Time expressed in multiples of measuring time at which a standstill is detected.
STIME 1…9 times measuring time, default = 1
6.3.3.13 Parameter P45: Standstill Range
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 SRANGE MSB
Byte 3 SRANGE LSB
Limit to define the weight to be in standstill.
SRANGE 0…10.00d default = 1d
Value = d*100
6.3.3.14 Parameter P46: Standstill Timeout
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2
Byte 3 TIME
Number of measuring times during the standstill condition should be reached.
TIME 0…100 M (measuring times), default = 8
If the standstill timeout is exceeded by a function, which needed the standstill condition, the function is
aborted (Taring, set zero, calibrate, 'P'-command of the SMA-protocol). Additional the error code 31 and the
'CmdError' bit are set.
6.3.3.15 Parameter P47: Zeroset Range
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0
Byte 0
Byte 1
Byte 2 RANGE MSB
Byte 3 RANGE LSB
RANGE 0…500.00 d default = 50 d
Value = d*100

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5211 Series and is the answer not in the manual?

Sartorius PR 5211 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5211 Series
CategoryTransmitter
LanguageEnglish

Related product manuals