IPE100 indicator NU-IPE100-IO-USER-E-0212
44
15. OTHER FUNCTIONS
15.1 CALCULATOR
Through function 114 combinable with the desired key (<< F.KEYS >> step, TECHNICAL MANUAL.), one enables the
calculator function.
• ADDITION
• MULTIPLICATION
• SUBTRACTION
Procedure:
- Enter the first value using the numerical keyboard.
- Press F1 to add, F2 to multiply, F3 to subtract the value.
Terminate the operation:
- TYPE the second value and PRESS ENTER: the result will be shown for a few seconds on the display.
- TYPE the second value and PRESS TARE: the result will be shown for a few seconds on the display and will be added
(if an addition and multiplication) or subtracted (if subtraction) to the current tare value.
- PRESS 2nd F: the NET weight on the scale will be used as a second value and the result will be shown for a few
seconds on the display.
In the end the printing of the “CALCULATOR” will be executed anyway (See section 14 PRINTOUTS).
To disable the calculator function, press the C key.
HELP FUNCTION
By pressing at length the . /HELP key it is possible to see the list of the keys used in the calculator function.
If one wants to scroll the list of the keys in manual mode, it is possible to use the arrow keys (F6 and F7 ).
15.2 DISPLAY OF NET WEIGHT WITH SENSITIVITY X 10 (for testing use during calibration)
By pressing at length the F2 key one will view the net weight with sensitivity times 10 (press again at length the F2 key to
return to the normal visualisation).
NOTES:
- With approved indicator, the visualisation lasts about 5 seconds, after which it’s disabled.
- The printout can only be done when the indicator has the standard sensitivity.
-
15.3 SETTING DATE / TIME
By pressing at length the F3 key ‘it’s possible to quickly access the setting from the date / time of the instrument:
DAY ENTER MONTH ENTER YEAR ENTER HOUR ENTER MINUTES ENTER.
15.4 SET-POINT FUNCTION
The indicator is fitted with a series of 4 signal outputs used for various types of operation, besides, through the I/O
expansion board (in the IO version) it is possible to use another 12.
In the SEtuP >>outPut >> r.ConF menu (TECHNICAL MANUAL) it is possible to set the characteristics of each output
which will be used: normally open or normally closed; upon direct control or weight stability, with hysteresis or without, the
functioning mode.
It is also possible to set whether the enabling of the output excludes the check on the previous outputs (see the SEtuP >>
outPut >> r.ModE parameter).