DEFIGARD/PHYSIOGARD Touch 7
12.6 Device Configuration
12.6.1 General configuration
Parameter Values Description
Notch filter
•None*
•50Hz
•60Hz
This option shall be activated if artefacts are detected on ECG signals
when the device is plugged to the mains. The notch filter must be
chosen according the location.
50Hz: Europe, Africa, Middle- East (except Saudi Arabia), Asia-
Pacific (except Japan, Taiwan and Philippines), Australia
60Hz: The American continent (except Chile, Argentina, Uruguay,
Paraguay, Bolivia, French Guyana)
Boot mode by
default
AED
Monitoring*
Manual def
Sets the mode your device shall start in when pressing the On/Off
button
Monitoring display
mode
Monitoring no curve
Advanced monitoring
Basic monitoring
12 leads ECG
Critical care
Sets the desired default view in monitoring.
Monitoring without curve: No curves are displayed by default, only big
monitoring values
Default heart rate
source
•Auto*
•Defi ECG
•Pleth
Sets the behaviour of the HR parameter box between following
possibilities:
Auto: The device automatically detects the HR source with
predefined priorities level. DEFI higher than ECG higher than SpO2
(pulse)
Defi: Always force the HR calculation on DEFI
ECG: Always force the HR calculation on ECG lead II SpO2: Always
force the HR calculation on SpO2 (pulse)
Audio pause at start
•2min*
Off
When this option is activated, the device will always remain silent for 2
minutes at start, even if an alarm occurs
Periodic test
frequency
•Daily*
Weekly
The device wakes up by default weekly to perform a self test. Is is
possible to set a daily test.
Hereafter, the details of content of the automatic and manual self test
Time of test 12
This parameter specifies when the device will automatically wake up to
perform its self test. This field must be specified in hours in 24 hours
format.
Always specify time in HH and not in HH:mm
Example: 13 refers to 1PM Example: 13:30 is not allowed
Technician
password
schiller Sets the password that will be asked to enter the Control Panel
Confirmation
required on mode
change
•True*
•False
If this option is activated, the device will ask a confirmation in following
situations
- Switch from AED to monitor
- Switch from AED to manual def
- Switch from monitor to manual def
Enable printer
•True*
•False
Activating this option will allow to pair a Bluetooth printer and will make
the printer button available
Curve thickness
0.5mm*
•0.7mm
•0.9mm
Sets the thickness of the curves printed on the Bluetooth printer.