Page 153
Technical Data 12
User guide Device Configuration 12.6
Art. no.: 0-48-0227 Rev.: g
DEFIGARD/PHYSIOGARD Touch 7
12.6.16 Transmission
12.6.17 Ethernet
Parameter Values Description
Automatic R-ECG
transmission
False*
•True
If this option is activated, the device will automatically send over the
media selected in configuration Transmission media during
intervention to the address specified for SEMA Server
There will be no preview of the resting ECG on the device, after
acquisition, it will be automatically sent
Automatic
intervention data
transmission
When device plugged on
docking station
At device power off
Off*
The device is capable of transmitting automatically its memory after an
intervention using the media selected in Transmission media by
automatic wake up, with the following scenario:
- Never (Off)
- 10 minutes after the device has been shutted down (After device
power off)
- When a external power supply is detected (When device plugged on
docking station)
Transmission media
during intervention
•Wi-Fi*
•GSM/3G
USB/Ethernet
The selected media will always be selected by default at manual start
of the device.
This media can always be changed by the user during the intervention
Transmission media
by automatic wake
up
•Wi-Fi*
•GSM/3G
USB/Ethernet
The selected media will always be selected by default during automatic
wake up of the device (automatic intervention data transmission,
automatic self test results transmission...)
Parameter Values Description
Ethernet ping serve 8.8.8.8
Sets the IP that will be used to test the connectivity of the device over
Ethernet. Usually, it's recommended to use either common DNS
server(e.g.
8.8.8.8), or the public SEMA Server IP address