Technical Data 12
User guide Device Configuration 12.6
Art. no.: 0-48-0227 Rev.: g
DEFIGARD/PHYSIOGARD Touch 7
12.6.3 Defibrillator
12.6.4 Display
Parameter Values Description
Default energy for
adults in manual def.
2, 4, 8, 15, 30, 50, 70, 90,
120, 150*, 200 Joule
Sets the energy which will be displayed by default when entering
manual defibrillation mode in adult mode
Default energy for
children in manual def.
2, 4, 8, 15, 30, 50*, 70, 90,
Joule
Sets the energy which will be displayed by default when entering
manual defibrillation mode in child mode
Sync. after sync shock
•True
False*
If this option is activated, the device remains in sync. mode after
a synchronized shock in manual defibrillation
First shock for adults
2, 4, 8, 15, 30, 50, 70, 90,
120, 150, 200 Joule
Sets the energy which will be delivered for the first shock in AED
in adult mode
Second shock for adults
2, 4, 8, 15, 30, 50, 70, 90,
120, 150, 200 Joule
Sets the energy which will be delivered for the second shock in
AED in adult mode
Third shock for adults
2, 4, 8, 15, 30, 50, 70, 90,
120, 150, 200 Joule
Sets the energy which will be delivered for the third shock in AED
in adult mode
First shock for children
2, 4, 8, 15, 30, 50*, 70, 90,
Joule
Sets the energy which will be delivered for the first shock in AED
in child mode
Second shock for
children
2, 4, 8, 15, 30, 50*, 70, 90,
Joule
Sets the energy which will be delivered for the second shock in
AED in child mode
Third shock for children
2, 4, 8, 15, 30, 50*, 70, 90,
Joule
Sets the energy which will be delivered for the third shock in AED
in child mode
Parameter Values Description
Anteriority Analysis
False*
•True
Enables the analysis with anteriotity (part of the signal analysis is done
on the signal before the device mentions the start of the analysis).
Analysis key
•False
True*
If this option is activated, the analysis key will be displayed to allow
additional manual analysis in AED mode.
Alarm on Vfib or
Vtach detection
•False*
True*
High priority alarm will be issued if Vfib or Vtach is detected during
monitoring on both ECG cable and Defi pads.
Precognitive
Analysis
False*
•True
Not used at the moment, has no impact on the unit.