5 Defibrillation
5.7 Defibrillator Technical Messages
Page 98
Art. no.: 0-48-0227 Rev.: g
DEFIGARD/PHYSIOGARD Touch 7
5.7 Defibrillator Technical Messages
Alarm Code Cause Remedy
Defibrillator inoperative
T.ECG11
T.DEFI01
T.DEFI02
T.DEFI03
T.DEFI04
T.DEFI05
T.DEFI06
T.DEFI07
T.DEFI10
T.DEFI13
T.DEFI14
T.DEFI15
CPU peripheral board defective
Defi board defective
Contact technical service
Incorrect electrodes
T.DEFI08
T.DEFI12
Internal discharge while releasing
shock “on impedance higher than
250 ohm” or if the current during the
shock is 0 or above 105 A, because
of poorly placed electrodes
Check electrodes, if necessary re-
apply electrodes
Internal discharge duration
too long
T.DEFI09
A shock that is not delivered within
the specified duration causes an in-
ternal discharge.
Do not exceed the time of 20 sec-
onds till releasing the shock
Contact technical service
Internal discharge smaller
energy
T.DEFI11
Internal discharge because of select-
ing lower energy after the device
charged the selected higher energy
Normal safety discharge