4 Monitoring
4.3 Operator-Defined Alarm Thresholds
Page 42
Art. no.: 0-48-0227 Rev.: g
DEFIGARD/PHYSIOGARD Touch 7
NIBP SYS [mmHg] Pat. value Wide limits Narrow Limits
Range SYS: 30-255 (50-200) mmHg Low High Low High
<90 -20 +35 -10 +25
90-114 -20 +35 -10 +25
115-140 -25 +35 -10 +20
>140 -25 +35 -10 +20
Allowed values [30-245] [30-245] [30-245] [30-245]
NIBP DIA [mmHg] Pat. value Wide limits Narrow Limits
Range DIA: 15-220 (20-150) mmHg Low High Low High
<65 -15 +25 -10 +25
65-90 -15 +15 -15 +10
>90 -15 +15 -15 +10
Allowed values [12-210] [12-210] [12-210] [12-210]
NIBP MAP [mmHg] Pat. value Wide limits Narrow Limits
Range MAP: 15-223 (20-235) mmHg - - - -
IBP SYS [mmHg] Pat. value Wide limits Narrow Limits
Range SYS: 30-255 (50-200) mmHg Low High Low High
<90 -20 +35 -10 +25
90-114 -20 +35 -10 +25
115-140 -25 +35 -10 +20
>140 -25 +35 -10 +20
Allowed values [30-245] [30-245] [30-245] [30-245]
IBP DIA [mmHg] Pat. value Wide limits Narrow Limits
Range DIA: 15-220 (20-150) mmHg Low High Low High
<65 -15 +25 -10 +25
65-90 -15 +15 -15 +10
>90 -15 +15 -15 +10
Allowed values [12-210] [12-210] [12-210] [12-210]
IBP MAP [mmHg] Pat. value Wide limits Narrow Limits
Range MAP: 15-223 (20-235) mmHg - - - -